What does kto stand for?
Keep Tube Open. When you want to stop the feeding for more than 30 minutes, KTO will periodically pump a medically insignificant amount to prevent the feed from clogging in the set.
Utbd del alarm keeps showing
It's actually "VTBD del" which means the "Volume To Be Delivered" has been set and the pump has delivered that much food. Pause the pump (Start/Hold button), then press and hold the VTBD button (below Start/Hold) for 5 seconds. That clears the VTBD.
How can I adjust the
You can't. It's a safety issue. If the pump turned off automatically the operator wouldn't know if feeding was completed successfully or if the pump lost power.
Is there a way to prime a ekangaroo (ref 773656)
Yes, The piece that you load into the pump is round and the center white piece will spin in a circle. That piece is a valve that will turn and let liquid pass through. Fill your bag with formula or water and then use a knife or screwdriver to turn that plastic white piece. Turn it slowly until you see the formula pass through it. Be careful though as the formula will go pretty quick. Turn the plastic circle piece back to original position to stop the flow.
Hope this helps. Good Luck
My daughter uses a joey at night for feeding and
How old is your daughter?? Is she able to let you know if she aspirates on her feed?? i recommend hand feeding for safety. Her stomache could be full and causing the back flow. Her feedings may have to be reduced/spread out.
No set
There may be a problem with the drip sensor.
If you have primed the pump set and place the drip chamber in the cutout area in the case and
you still get the same error, then return the pump to the company that provided it for you so it can be serviced by a trained technician.
Not finding what you are looking for?