My pocket scales reads - oue2 when i turn it on .what does this mean
Special Calibration- If the display shows [OUE2] after turning on: 1. Turn OFF the scale. 2. Press ON/OFF key briefly then immediatley press and hold the MODE and TARE buttons simultaneously until [CAL] is displayed. Release the MODE and TARE buttons and a random internal statement number will be displayed. 3. Press the MODE key once and the display will show [2Ero]. 4. Place a 200g weight on the center of the platform. Wait 2 seconds then press MODE key again. The LCD will display [PASS] then [200.0]. 5. Remove the weight and turn off the scale. Special Calibration is complete.
My digital pocket scales read zero even with weight on them?
- Clean the Scale. Make sure that the pocket scale is completely clean. ...
- Reset Scale to Zero. You want to reset the scale so that it is at zero. ...
- Find the Calibration Weight. ...
- Find Nickels for Great Substitute Weights. ...
- Calibrate. ...
- Check the Calibration.
My digital pocket scales say over
You did not post the make and model but they have been knocked out of calibration. There may be no way to reset them. When they are turned on there is a sensor in them that is queried by the processor and there is an expected response. The response is wrong and you get the error that you stated.