I have a microlife AG,9443 Widnaut to mesure arterial tension What is the error 3

2 Related Answers

Anonymous

  • 1 Answer
  • Posted on Dec 04, 2010

SOURCE: what does error 3 on my microlife blood pressure

I have digital blood pressure monitor Model 302 made in japan. It is not working. when start shoes error 3. Please advise

Ad

Erik Walters

  • 155 Answers
  • Posted on Jan 10, 2015

SOURCE: what does error 3 on my microlife blood pressure monitor

Repeat measurement keeping arm still. If inflation of the cuff takes too long, the cuff is not correctly seated or the hose connection's not tight. Check connections and repeat

Add Your Answer

×

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

0helpful
1answer

I need downloads drive to connect my Microlife BP 3AG1 to my laptop

This model DOES NOT connect to a computer. download the manual here. Press CTRL+F, type 'computer'. Nothing!

BP 3AG1 - Microlife AG

https://www.microlife.com > support > blood-pressure
BP 3AG1 Blood pressure monitor Model no.: BP 3AG1. Dimensions: 140 x 120 x 70 mm. Weight: 400g (including batteries) Voltage source: 4 x 1.5 V alkaline batteries; size AA or mains adapter DC 6V, 600 mA (optional)
0helpful
3answers

I have microlife model BP3GX15A blood pressure monitor and it will not work, It keeps giving err number 3. I have replaced the batteries twice and it still doesnt work

For most Microlife blood pressure monitors, the error 3 indicates that the unit cannot reach the proper pressure in the cuff to start the test. If replacing the batteries doesn't work, check that the cuff is properly plugged into the socket on the monitor. Next you may need to replace the cuff. There may be a leak in the cuff or the hose between the cuff and the socket. Microlife sells replacement blood pressure cuffs here: https://www.microlife.com/consumer-products/blood-pressure/cuffs-and-supplies . These cuffs fit all of their devices except for the wrist versions.

I hope this helps.

Cindy Wells
0helpful
1answer

What does error 3 on my microlife blood pressure monitor

Repeat measurement keeping arm still. If inflation of the cuff takes too long, the cuff is not correctly seated or the hose connection's not tight. Check connections and repeat
0helpful
2answers

Error code 3 on a Blood Presure unit Microlife Bp A100 Plus what can i do to fix this problem ?

i did have the same problem..and i resolved it..but before i give you the solution..do you hear the pump is working or not when you switch the device on? if not then one of the transistors are gone off. see my channel "tareq el-hasan" on youtube..i will upload the video very shortly..best luck
0helpful
1answer

What does error 2 on my microlife blood pressure monitor mean?

ERROR 2 IN MICROLIFE PRESSURE MONITOR MEANS : Unnatural pressure impulses influence the measurement result. Reason: The arm was moved during the measurement (Artefact).
0helpful
2answers
34helpful
6answers

What does error 3 on my microlife blood pressure monitor mean?

I have digital blood pressure monitor Model 302 made in japan. It is not working. when start shoes error 3. Please advise
Not finding what you are looking for?

370 views

Ask a Question

Usually answered in minutes!

Top Microlife Health & Beauty Experts

Rob Hill
Rob Hill

Level 3 Expert

1480 Answers

ADMIN Andrew
ADMIN Andrew

Level 3 Expert

66835 Answers

ADMIN Eric
ADMIN Eric

Level 3 Expert

39363 Answers

Are you a Microlife Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...