HP 116 electronic fetal monitor (same as Corometrics 116) error code HLP 3
Reply from Boyd Campbell CBET, CRES
I have seen two different issues cause this error code. The most common is the DSP board and the second cause could also be the main processor. If you have another unit I would try to switch the DSP first. If you look at the unit from the front the DSP board is the one on the far right. Unfortunately I do not know of a way to troubleshoot that error down to the component level. If you want to give us a call we can assist you with the parts. Boyd Campbell CBET, CRES Southeatern Biomedical 1-888-310-7322
5/13/2017 4:09:03 PM •
Medical
•
2,311 views • 0 helpful votes
Number 6 button flashing what is this
Please be more specific. What is the Make and model number of your machine? Are you still having this issue? If you restart from stop, exactly what happens? This info is needed for proper troubleshooting.