Troubleshooting
E22 - Optical
reading is out
of range
A Sample or Reference
channel reading taken
during an Air
measurement is too
high.
1. Discard the sample.
2. Run an optical test.
3. If you are unable to run an
optical test, contact your local
technical support provider.
E23 -
Excessive
noise on the
Sample
channel
The standard deviation
of the 16 readings at the
Sample channel for Dark
or Air reading is too
large.
1. Discard the sample.
2. Run an optical test.
3. If you are unable to run an
optical test, contact your local
technical support provider.
E24 -
Excessive
noise on the
Reference
channel
The standard deviation
of the 16 readings at the
Reference channel for
Dark, Air or Sample
reading is too large.
1. Discard the sample.
2. Run an optical test.
3. If you are unable to run an
optical test, contact your local
technical support provider.
E26 -
Excessive
noise in sample
reading
The standard deviation
of the 16 readings at the
Sample channel during a
reading taken is too
large in Sample read
position.
1. Discard the sample.
2. Run an optical test.
3. If you are unable to run an
optical test, contact your local
technical support provider.
E27 -
Excessive
Lamp Drift
The change in mean
signal between
successive Air readings
is too large at either the
Sample channel or the
Reference channel.
1. Discard the sample.
2. Run an optical test.
3. If you are unable to run an
optical test, contact your local
technical support provider.
E30 - Thermal
control system
error - low
The temperature
measured by one of the
cartridge holder
thermistors is ?2?C.
1. Discard the sample.
2. Restart the system.
3. If you are unable to perform a
successful restart, contact
your local technical support
provider.
E
×