DigiWeigh 100 X 0.01 Gram Digital Jewelry Pocket Scale Logo
Posted on Nov 09, 2010
Answered by a Fixya Expert

Trustworthy Expert Solutions

At Fixya.com, our trusted experts are meticulously vetted and possess extensive experience in their respective fields. Backed by a community of knowledgeable professionals, our platform ensures that the solutions provided are thoroughly researched and validated.

View Our Top Experts

My scale is frozen displaying f--h how do I fix this? I tried holding mode and tare and turning it onon and that doesn't work any other suggestions?

1 Answer

Doctor PC

Level 3:

An expert who has achieved level 3 by getting 1000 points

Superstar:

An expert that got 20 achievements.

All-Star:

An expert that got 10 achievements.

MVP:

An expert that got 5 achievements.

  • Master 7,733 Answers
  • Posted on Nov 12, 2010
Doctor PC
Master
Level 3:

An expert who has achieved level 3 by getting 1000 points

Superstar:

An expert that got 20 achievements.

All-Star:

An expert that got 10 achievements.

MVP:

An expert that got 5 achievements.

Joined: Oct 01, 2010
Answers
7733
Questions
0
Helped
1864810
Points
21487

Reset it!

Turn off the scale. Press and hold the MODE button and turn the scale on. Release the MODE button you will see the scale's internal count.
Press and hold the TARE button then press (don't hold) the MODE button you will see the full capacity is changing. Keep pressing the MODE while holding the TARE button until you see 100.00 on the screen. The reset is completed. You may need to recalibrate the scale after the reset.

Add Your Answer

×

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

0helpful
1answer

My pocket scales reads - oue2 when i turn it on .what does this mean

Special Calibration- If the display shows [OUE2] after turning on: 1. Turn OFF the scale. 2. Press ON/OFF key briefly then immediatley press and hold the MODE and TARE buttons simultaneously until [CAL] is displayed. Release the MODE and TARE buttons and a random internal statement number will be displayed. 3. Press the MODE key once and the display will show [2Ero]. 4. Place a 200g weight on the center of the platform. Wait 2 seconds then press MODE key again. The LCD will display [PASS] then [200.0]. 5. Remove the weight and turn off the scale. Special Calibration is complete.
0helpful
1answer

What does oue2 mean on the scale

When you see( 0UE2 )error message on your scale display after turning on , the following method is used for restore your precious balance. This method is called Special Calibration-
1. Turn OFF the scale.
2. Press ON/OFF key briefly then immediatley press and hold the MODE and TARE buttons simultaneously until [CAL] is displayed. Release the MODE and TARE buttons and a random internal statement number will be displayed.
3. Press the MODE key once and the display will show [2Ero].
4. Place a 200g weight on the center of the platform. Wait 2 seconds then press MODE key again. The LCD will display [PASS] then [200.0]. 5. Remove the weight and turn off the scale.
Special Calibration is complete.

With best regards

HUK
CIGTL-Pak gemological Lab
1helpful
1answer

I bought a pierre cardin digital kitchen scale YHC1407C online including battery.is the battery inserted properly when u packed or should we insert it.the machine is not responding on pressing TARE...

If the scale is ''on'' then the battery is in correctly. ''Tare'' is used for making the scale read ''zero'' when a ''holder'' (such as a cup or plate) is placed on the scale and you want the scale to be at ''zero'' with that item on top of the scale. With nothing on the scale, the tare button will not do anything. Also, you have to hold down the tare button until the scale actually displays ''zero.'' (Perhaps you are just ''pressing the button'' and releasing it? You need to press and _hold_ the button down until the scale actually reads zero).
Jan 01, 2015 • Cameras
0helpful
1answer

How can you fix error on screen can it be reset

It often just needs recalibrating. Try turning the scale on while holding the mode/tare button with nothing on the scale. Then press the tare button again. It should show a calibration weight which you should put onto the scale. Then it automatically calibrates.
0helpful
1answer

I have a DigitZ DZ3-600 and when I turn it on the word FULL in the display screen . What can I do to clear this and start using it?

My search is telling me that this is a scale. Is this correct? If so - this is the manual. \'FULL\' is not an error, but this is the correct model. If this is not a scale, please write again and provide more information:



The manual is short and PASTED after link to same here:



http://digitzscales.com/index.php?option=com_content&view=frontpage



Digitz DZ3-600 Manual




I. Operation


1. Place the scale on a flat, level surface and press the ON/OFF key to turn the scale on.

2. Once the start-up process completes and the scale goes to zero you are ready to weigh.

3. Place the quantity to be weighed onto the weighing platform.

4. Wait a couple seconds for the display to stabilize before taking a weight reading.



II. Changing the Unit of Measure


1. Press the MODE key to change the scales displayed unit of measure.

2. If the scale is turned off then back on, it will default to the last unit used.

• DZ3-100 Units: g (gram) oz (ounce) ct (carat) gn (grain)

• DZ3-600 Units: g (gram) oz (ounce) ozt (troy ounce) dwt (pennyweight)



III. Using the Tare Feature


You can use this scale\'s tare feature to subtract the weight of an empty container or vessel for net weight determination. NOTE: The weight of the container will subtract from the available capacity.

1. Place the empty container or vessel on the weighing platform.

2. Press the TARE key once. The display will then reset to zero. If the container is removed, it\'s weight will be displayed as a negative number.

3. You may now fill the container with the quantity to be weighed. The display will show the net weight of the contents.

4. To return to normal weight display (gross weight), remove any items from the weighing platform and

press the TARE key again to re-zero the scale.



IV. Calibration / Adjustment


1. Turn the scale on by pressing the ON/OFF key.

2. Wait for the display to show "0.0g" or "0.00g" then press and hold the MODE key until the display shows "CAL", then release.

3. Press the MODE key again. The display will flash "CAL" followed by the required calibration weight.

4. Place the required calibration weight on the center of the platform. The display will show "PASS" then return to normal weighing mode. Calibration is complete! You may now turn the scale off.



V. Error Codes


1. OUTZ - Zero range has shifted. Perform calibration adjustment. If the problem persists, the load cell may be permanently damaged due to overloading.



2. 88888 - Maximum capacity exceeded. Remove the extra weight to avoid damaging the load cell.



3. L - Batteries are low. Replace the batteries.



VI. Specifications


DZ3-100 DZ3-600

Max. Capacity 100g 600g

Readability 0.01g 0.1g

Min. Weight 0.03g 0.2g

Power Two (2) AAA Batteries

Dimensions 3.0 x 5.0 x 0.8 in.

Platform Dimensions 2.5 x 3.0 in.



For warranty and support, please email

[email protected]
0helpful
1answer

Pocket scales read hhh turn

Try this - While it is off, press and hold both the mode and the tare* keys at the same time and press the On/Off key - Release the keys and observe the display - cycle through a series of numbers. Press the mode key once and observe the display "CAL 0" "100.00" then similar numbers as before. This step calibrates
Check here - http://wiki.answers.com/Q/How_do_you_calibrate_digi_weigh_scale_DW-100AX_or_fix_errors_such_as_F--H_or_ERR?#slide=1
17helpful
2answers

When i turn on my digiweigh dw 100ax scales F--H appears on the screen how do i fix that and calibrate the scales when done

I thought the scale was just completely broken, I was able to fix by doing the following.

1. turn unit off
2. hold down the on/off/tare AND the MODE button.
3. The unit will come up with a bunch of numbers on the screen. Release the buttons
4. Press MODE once, you will see "100" flash on the screen. if you don't press the button a second time.
5. place 100g of tare weights on the scale.
6. Press MODE button again. After this my scale said PASSED and shut off.
7. remove tare weights and turn device on. It should now register properly.

In my case. the scale was misregistering the 50g weights as 75. I shut it off. I pulled up ever so slightly on the scale plate on the hinge side of the plate are repeated steps 1-7.

I have no idea why they don't have a 3 part calibration on these.

It is working like a charm now!



12helpful
2answers

Scale displays OUT2

Hi,
you need just to recalibrate it:
The procedure is standard for semi-all models of pocket scales, but the way how to get to calibration mode differs from model to model (or vendor to vendor).
In brief:
1.While turn-on get to the calibration mode (by pressing and holding the ON button, or holding TARE while turinig on) You should hold this "magic" combination until the LCD will show a number (large number - it was more than 2000 in my case). This number is a raw data from the sensor.
2.Now release the "magic" button.
3.Press it one more time, the display will show 2Ero and then 100.0
4.Now put calibrated weight of 100 grams on the scale and wait 3 seconds.
5.Press "magic" button again.The display will show CAL and then PASS.
6.Now remove the weight and turn off the scale.
7.Thats all - now you can use your scale as usually.
15helpful
2answers

F--H APPEARS WHEN TURNED ON AND FREEZES

that dosent work in the case of an f--h display at least in my case i think the f--h stands for f**king hopeless
Not finding what you are looking for?

4,101 views

Ask a Question

Usually answered in minutes!

Top DigiWeigh Measuring Tools & Sensors Experts

Brad Brown

Level 3 Expert

19187 Answers

Steve Pack
Steve Pack

Level 3 Expert

7128 Answers

ADMIN Andrew
ADMIN Andrew

Level 3 Expert

66957 Answers

Are you a DigiWeigh Measuring Tool and Sensor Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...