I have digital blood pressure monitor Model 302 made in japan. It is not working. when start shoes error 3. Please advise
I have a digital pressure monitor model :BP38TO-A made in Switzerland, It is not working. When start shoes Error 3. Please advise.
A full answer can be found in the user manual, or on the official website of the brand.
Buy silagra 100mg online Online is used for men, High blood pressure and other conditions. Silagra 100mgTablet may also be used for purposes not listed in this medication guide. Order Silagra 100mg Online From Trusted Online Pharmacy Medypharma . Best Price In USA.
Three main causes for Err 3:
1. Air leakage===> check cuff, air outlet socket
2. solenoid not functioning properly===> check solenoid and/or transistor Q3 on main board
3. motor is not functioning===> check the motor and /or transistor Q4 on the main board (see my youtube channel tareq elhasan)
I have a Tensoval comfort I bought it a very long time ago.
now it says Error 3
today to my surprise it worked.but when i used it again it said the same error 3..Is there something i can do to fix it?
George 71 yo Grecce
27,545 views
Usually answered in minutes!
u have a air leak,chk hose and cuff
An adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if neces- sary. Repeat the measurement.
×