Anonymous Posted on Dec 19, 2013

Mitutoyo e671 cmm - Measuring Tools & Sensors

Add Your Answer

×

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

0helpful
1answer

Mitutoyo HDS-H24"C the NUMBERS JUMPING aROUND.

Send it in for repair and calibration

Repair, Calibrate or Request Field Service

Mitutoyo https://www.mitutoyo.com > Contact Us
Our highly skilled technicians provide fast, accurate and dependable service to keep your production operation running smoothly.
0helpful
1answer

Mitutoyo error code E552?

Error code E552 is usually caused by a communication issue between the CMM and the computer. You can try restarting the CMM and the computer, and ensuring that all cables are properly connected. If the issue persists, you may need to take your CMM to a certified Mitutoyo service center for further diagnosis and repair.
0helpful
1answer

Mitutoyo error code E940?

Mitutoyo is a brand that produces precision measuring instruments such as micrometers, calipers, and height gauges. An error code E940 on a Mitutoyo instrument typically indicates a problem with the instrument's communication interface. This can be caused by a variety of issues, such as a problem with the instrument's cable or software, or a problem with the computer or device to which the instrument is connected.
To troubleshoot the issue, you should first check the connections between the instrument and the computer or device to ensure that they are properly connected and secure. If the connections appear to be fine, you may need to update the software on the instrument and computer or device to resolve the issue.
0helpful
1answer
0helpful
1answer
0helpful
1answer

Mitutoyo micrometer/caliper. How do I install new glass/plastic dial cover on 6 Mitutoyo micromonitor / calipher.

Repair, Calibrate or Request Field Service

Mitutoyo https://www.mitutoyo.com > Contact Us
Whether you need immediate repair service or A2LA Accredited, NIST-traceable calibration of your measuring instruments, our highly skilled technicians provide ...

Mitutoyo Repair and Calibration

Indicator Repair https://www.indicatorrepair.com > manufacturers > mit...
Indicator Repair repairs, restores, refurbishes and calibrates all makes and models of Mitutoyo indicators, micrometers, gages, calipers and levels.

..
0helpful
1answer

I have an error on my mitutoyo CMM showing E650 on Joy stick

Verifica la presi?n de aire en la v?lvula de presi?n en la parte frontal de la cmm
1helpful
2answers

How to correct an e671 error message

I realise this question is old, but you never know, might help someone in the future. This particular error is that the probe is not correctly seated, usually the magnets have become slightly loose. The solution is to take the probe head off the cmm, and push the magnets against your surface table or suitable flat surface, this should push the magnets into their correct position. If this doesn't help, it may be that you've managed to break that particular part, if you have a probe rack, try a different probe. Lastly, if you crashed the CMM by hitting something, ensure you didn't bend your probe in the process (and the ruby is still there!), this doesn't help the error code fix, but worth checking as the cmm crashing into something at speed is the most likely cause of this E671 error.
Not finding what you are looking for?

905 views

Ask a Question

Usually answered in minutes!

Top Measuring Tools & Sensors Experts

Brad Brown

Level 3 Expert

19187 Answers

Steve Pack
Steve Pack

Level 3 Expert

7128 Answers

ADMIN Andrew
ADMIN Andrew

Level 3 Expert

66949 Answers

Are you a Measuring Tool and Sensor Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...